Multiscan 3D ultrasound stitching for large-volume abdominal imaging: a phantom and prospective feasibility study in healthy volunteers and selected patients

Blank, Valentin, Schwabe, Paula, Wiegand, Johannes, Karlas, Thomas
Springer Nature · 2026-11-23 · DOI 10.1186/s12880-026-03031-6

Abstract

Background Conventional abdominal ultrasound is limited by operator-dependent acquisition, non-standardized documentation, and a restricted field of view. Tomographic three-dimensional ultrasound combined with multiscan stitching may overcome some of these limitations by generating structured reviewable large-volume datasets that allow retrospective multiplanar review. This pilot study evaluated the technical feasibility of magnetically tracked multiscan three-dimensional ultrasound (3DUS) stitching for large-volume abdominal imaging. Methods A multiscan stitching algorithm was first assessed qualitatively under controlled conditions using standardized ultrasound phantoms to evaluate volume reconstruction, seam continuity, and artifact formation. The method was then prospectively applied in 22 healthy volunteers and 22 patients with focal or chronic parenchymal liver disease in a single-centre observational within-participant comparison. Conventional ultrasound served as the clinical comparator for image quality and completeness of organ depiction, paired ordinal comparisons were exploratory and post hoc. Additional phantom and proof-of-concept torso examinations were unscored. Results In phantom experiments, adjacent three-dimensional ultrasound volumes could be merged with visually preserved margin continuity under standardized acquisition conditions and sufficient nominal overlap. In healthy volunteers, stitched Scan 3 yielded a median image-quality score of 3 (IQR 2–3) versus 4 (4–4) for conventional ultrasound and a median organ-coverage score of 20 (18–20.75) versus 27 (26–27.75), respectively (both Holm-adjusted P < 0.001). In patients, non-stitched Scan 6 yielded an image-quality score of 4 (3–4) versus 3.5 (3–4, adjusted P = 0.289), whereas organ coverage remained lower at 17 (16.25–17) versus 27 (24.25–27, adjusted P < 0.001). Focal lesions and features of chronic liver disease were visible in reconstructed datasets, but diagnostic accuracy was not assessed. Stitching-related artifacts, respiratory motion, transducer pressure, and tracking-related limitations were identified as relevant technical challenges. Conclusion Multiscan tomographic three-dimensional ultrasound enabled large-volume abdominal ultrasound recording, with at least one analysable 3DUS dataset obtained for each of the 44 enrolled participants under favourable examination conditions. The stored datasets allowed retrospective multiplanar review, however, the fixed-reader design did not establish examiner independence or interobserver reliability. The findings remain exploratory and system-specific. Quantitative validation of geometric accuracy, diagnostic performance, reproducibility, and robustness in larger and more heterogeneous patient cohorts is required before broader clinical implementation. Clinical trial registration The non-interventional study was registered in the German Clinical Trials Register (DRKS00016736) on 27 May 2019, before the first participant was enrolled on 28 May 2019. The registry planned 50 participants, whereas 44 participants were included in the present analysis.

Read at publisher →