Analytical Validation of β‐Hydroxybutyrate Quantification in Feline Serum with the <scp>BT3500</scp> Analyzer

Alessandro Tirolo, Maria Chiara Sabetti, Alicia Maria Carrillo Heredero, Francesca Fidanzio, Andrea Corsini
Wiley · 2026-11-11 · DOI 10.1111/vcp.70163

Аннотация

ABSTRACT Background Quantitative measurement of serum β‐hydroxybutyrate (β‐OHB) is essential for detecting and monitoring ketosis and ketoacidosis in diabetic cats, particularly in the era of sodium–glucose cotransporter two (SGLT2) inhibitors, where euglycemic ketoacidosis may occur. The BT3500 analyzer has not been analytically validated for β‐OHB quantification in feline serum. Objectives Analytically validate the Randox Ranbut enzymatic β‐OHB assay on feline serum using the BT3500 chemistry analyzer, including evaluation of sample stability under refrigerated and frozen storage conditions. Methods Residual feline serum samples and pooled sera were used to assess intra‐assay and inter‐assay precision, spike–recovery accuracy across five concentration increments, linearity under serial dilution, and analytical sensitivity (LoB, LoD, LoQ). Short‐term stability was evaluated at 4°C for 7 days and long‐term stability at −20°C at predefined time points up to 360 days. Results Intra‐assay coefficient of variation (CV%) ranged from 1.68% to 14.48% and inter‐assay CV% from 3.08% to 10.20%, with higher imprecision at low concentrations. Recoveries ranged from 110% to 122% (mean 114% ± 4%). Linearity was excellent (slope 1.012; R 2 ≥ 0.9998). LoB, LoD, and LoQ were 0.044, 0.091, and 0.150 mmol/L, respectively. β‐OHB was stable at 4°C up to 6 days and at −20°C up to 180 days, while decreases exceeding the acceptance limit were observed at 360 days. Conclusions The Randox β‐OHB assay on the BT3500 demonstrated acceptable analytical performance in feline serum and is suitable for routine clinical use; frozen storage beyond 180 days may impact result accuracy.

Читать у издателя →