Comparison of joint laxity measured using 60- and 80-N distraction forces and manual stress during robot-assisted total knee arthroplasty

Hiroshi Inui, Haruhiko Nakamura, Ryota Takei, Masaki Uchida, Kei Satou, Tetsu Yamashita, Risa Matsumoto
Springer Nature · 2026-11-10 · DOI 10.1186/s42836-026-00449-9

Аннотация

Abstract Background Appropriate intraoperative soft-tissue balance is essential for achieving successful outcomes following total knee arthroplasty (TKA). However, the optimal distraction force for evaluating joint laxity remains unclear. This study aimed to compare joint laxities measured using specific distraction forces with those obtained under manual varus-valgus stress and assess the interobserver variability. Methods A total of 55 bi-cruciate-stabilized TKAs were performed using a robotic system. Medial and lateral joint laxities were evaluated using a force-controlled tensioner device under a distraction force of 60 and 80 N at 10°, 30°, 60°, and 90° of flexion and under manual varus-valgus stress using the robotic system. To assess interobserver differences, all measurements were performed independently by the senior surgeon and one of the assistant surgeons. Results In the medial compartment, no statistically significant differences were detected between joint laxity under manual stress and that measured using a distraction force of 80 N at 10° and 30° of flexion. At 90° of flexion, no statistically significant difference was detected between JLR and JLT60N. In the lateral compartment, no statistically significant differences were detected between joint laxity under manual stress and that measured using 80 N at 10° and 90° of flexion, whereas joint laxity under manual stress was significantly greater than that measured using 80 N at 30° and 60° of flexion. Interobserver variability was greater in the lateral compartment than in the medial compartment, especially at higher flexion angles. Conclusions The differences between joint laxity measured under manual stress and that measured using a tensioner device varied depending on the compartment and knee flexion angle. The lateral compartment showed greater measurement variability, suggesting the need for careful assessment and standardization of intraoperative joint laxity evaluation during robot-assisted TKA.

Читать у издателя →